hi
have just got on of refactometers and it says to calerbrate it i usedistilled water is it ok to use ro water
hope you can help
garry
Yes Garry
Refractometers always read the SG at a water temperature of 20C regardless of the tank temperature and therefore to calibrate the refractometer properly the instrument must be at 20C during calibration and not the water.
When you then take a reading the REFRACTOMETER will always measure the SG as if the water is at 20C as the small amount of water that you add to the slide will have no impact on the temperature of the equipment and will soon change to the same temperature as the refractometer.
What ATC (auto temperature compensation) means is that once calibrated at 20C you can use the refractometer in higher ambient temperatures where the instrument is hotter or colder.
Once you get your reading you need to compensate for the temperature as most people quote an SG at 25C.
|
Temp. (°C) |
S.G. |
|
18 |
1.0268 |
|
19 |
1.0264 |
|
20 |
1.0262 |
|
21 |
1.026 |
|
22 |
1.0258 |
|
23 |
1.0256 |
|
24 |
1.0253 |
|
25 |
1.025 |
|
26 |
1.0246 |